Remote control commands
R&S
®
RTE
1168User Manual 1326.1032.02 ─ 20
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
18.9.2 Edge trigger
TRIGger<m>:EDGE:SLOPe.......................................................................................... 1168
TRIGger<m>:ANEDge:COUPling...................................................................................1168
TRIGger<m>:ANEDge:FILTer........................................................................................ 1169
TRIGger<m>:ANEDge:CUToff:HIGHpass........................................................................1169
TRIGger<m>:ANEDge:CUToff:LOWPass........................................................................1170
TRIGger<m>:ANEDge:GND..........................................................................................1170
TRIGger<m>:ANEDge:SLOPe.......................................................................................1170
TRIGger<m>:EDGE:SLOPe <Slope>
Defines the edge for the edge trigger event.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<Slope> POSitive | NEGative | EITHer
See Chapter 18.4.3, "Slope parameter", on page 1075.
*RST: POSitive
Usage: Asynchronous command
TRIGger<m>:ANEDge:COUPling <Coupling>
Sets the coupling for the external trigger signal.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Coupling> DC | DCLimit | AC
DC
Direct connection with 50 Ω termination, passes both DC and
AC components of the trigger signal.
DCLimit
Direct connection with 1 MΩ termination, passes both DC and
AC components of the trigger signal.
AC
Connection through DC capacitor, removes DC and very low-fre-
quency components.
Trigger