Remote control commands
R&S
®
RTE
1202User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
<n> 1..31
Indicates the trigger source: see TRIGger<m>:LEVel<n>:
NOISe[:STATe] on page 1201.
Parameters:
<HystMode> ABS | REL
ABS
The hysteresis is set in absolute values (voltage).
REL
The hysteresis is defined in relative values (div).
*RST: ABS
Usage: Asynchronous command
TRIGger<m>:LEVel<n>:NOISe:ABSolute <HystAbs>
Defines a range in absolute values around the trigger level. If the signal oscillates
inside this range and crosses the trigger level thereby, no trigger event occurs.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
<n> 1..31
Indicates the trigger source: see TRIGger<m>:LEVel<n>:
NOISe[:STATe] on page 1201.
Parameters:
<HystAbs> Range: 0 to The value corresponding to full division range.
The exact maximum value depends on the selected
vertical scale.
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
TRIGger<m>:LEVel<n>:NOISe:PERDivision <HystInDivs>