Remote control commands
R&S
®
RTE
1291User Manual 1326.1032.02 ─ 20
MEASurement<m>:STATistics:HBINs <HistogramBins>
Sets the number of bins - the number of vertical bars that build the histogram.
If MEASurement<m>:VERTical:CONT is ON, the instrument determines the number
of bins automatically based on the time base, the current measurements, and other
settings.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<HistogramBins> Range: 2 to 1000
Increment: 10
*RST: 1000
Usage: Asynchronous command
MEASurement<m>:STATistics:MODE <ResetMode>
Defines when the statistics for long term measurements are reset.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<ResetMode> TIME | WFMS | MEAS
TIME
Sets one long term measurement point after the time defined
using MEASurement<m>:STATistics:RTIMe.
WFMS - Waveforms
Sets one long term measurement point after a number of
acquired waveforms defined using MEASurement<m>:
STATistics:RCOunt.
MEAS
Sets one long term measurement point after a number of mea-
surement results.
*RST:
TIME
Usage: Asynchronous command
MEASurement<m>:STATistics:RCOunt <RstWfmCnt>
Defines the number of measured waveforms from which one point of the long term
measurement is created (reset of statistics).
Automatic measurements