Remote control commands
R&S
®
RTE
1308User Manual 1326.1032.02 ─ 20
MEASurement<m>:ONViolation:TRIGgerout <TriggerOutPulse>
Creates a trigger out pulse on limit or margin violation or on successful completion of
themeasurement.
Suffix:
<m>
.
1..10
Parameters:
<TriggerOutPulse> NOACtion | SUCCess | VIOLation
*RST: NOACtion
Usage: Asynchronous command
MEASurement<m>:ONViolation:RUNexec <RunExecutable>
Starts an external application if the command is set to VIOLation or SUCCess.
Use the following commands to set up the application:
●
EXECutable:NAME on page 1106
●
EXECutable:PARameter on page 1106
●
EXECutable:WDIRectory on page 1106
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<RunExecutable> NOACtion | SUCCess | VIOLation
See Chapter 18.4.5, "Event parameter", on page 1075
*RST: NOACtion
Usage: Asynchronous command
18.12.13 Reference levels
● General reference level settings......................................................................... 1308
● Automatic configuration.......................................................................................1310
● Manual configuration...........................................................................................1311
● Results................................................................................................................ 1314
18.12.13.1 General reference level settings
REFLevel<m>:LDETection............................................................................................1309
REFLevel<m>:LMODe................................................................................................. 1309
REFLevel<m>:RELative:MODE.....................................................................................1309
Automatic measurements