Remote control commands
R&S
®
RTE
1484User Manual 1326.1032.02 ─ 20
*RST: STBT
Usage: Asynchronous command
TRIGger<m>:UART:SOURce <Source>
Selects the transmitter or receiver line as trigger source.
Parameters:
<Source> TX | RX
*RST: TX
Usage: Asynchronous command
TRIGger<m>:UART:DPOPerator
<DataPosOperator>
Sets the operator for the data position. You can defined an exact position, or a position
range.
Parameters:
<DataPosOperator> EQUal | GETHan | INRange | RANGe
INRange = RANGe
*RST: GETHan
Usage: Asynchronous command
TRIGger<m>:UART:DPOSition <DataPosition>
Sets the number of words before the first word of interest. These offset words are
ignored.
Parameters:
<DataPosition> Number of words
Range: 0 to 32767
Increment: 1
*RST: 0
Usage: Asynchronous command
TRIGger<m>:UART:DPTO <DataPositionTo>
Defines the last word of interest, if TRIGger<m>:UART:DPOPerator defines a posi-
tion range.
Parameters:
<DataPositionTo> Range: 0 to 32767
Increment: 1
*RST: 0
Usage: Asynchronous command
Protocols