TRIGger<m>:CAN:SYMBolic:MSGValue <MessageName>
Sets the message to be triggered or searched for.
The setting is used by symbolic trigger and symbolic search. Thus, you always trigger
on and search for the same message if symbolic trigger and symbolic search is used at
the same time.
See also: SEARch:TRIGger:CAN:SYMBolic:MSGValue on page 1527
Parameters:
<MessageName> String with the symbolic message name
TRIGger<m>:CAN:SYMBolic:TSIGnals
<TriggerOnSignal>
Enables the trigger on a specific signal value that is part of the selected message.
Parameters:
<TriggerOnSignal> ON | OFF
*RST: OFF
TRIGger<m>:CAN:SYMBolic:SIGValue <SignalName>
Sets the signal name to be triggered or searched for.
The setting is used by symbolic trigger and symbolic search. Thus, you always trigger
on and search for the same signal if symbolic trigger and symbolic search is used at
the same time.
See also: SEARch:TRIGger:CAN:SYMBolic:SIGValue on page 1528
Parameters:
<SignalName> String with the signal name as defined in the DBC file.
TRIGger<m>:CAN:SYMBolic:DMAX <DataPatternTo>
Defines the maximum data value of the signal.
This value is required to specify a range if condition INRange | OORange is set with
TRIGger<m>:CAN:DCONdition.
Parameters:
<DataPatternTo> Decimal representation of the data pattern
Range: -100E+24 to 100E+24
Increment: 0.5
*RST: 1
TRIGger<m>:CAN:SYMBolic:DMIN <DataPattern>
Defines the minimum data value of the signal.
To set the condition, use TRIGger<m>:CAN:DCONdition.