Remote control commands
R&S
®
RTE
1541User Manual 1326.1032.02 ─ 20
BUS<m>:LIN:FRAMe<n>:DATA?
Returns the data bytes of the specified frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<Data> Comma-separated list of integer values (N, D1, D2,…, DN). N is
the number of bytes in the frame, and D1…DN are the values of
the bytes.
Example:
BUS:LIN:FRAMe4:DATA?
<-- 4,118,39,71,123
Usage: Query only
BUS<m>:LIN:FRAMe<n>:IDSTate?
Returns the identifier state of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<IdentifierState> OK | STERror | SPERror | PRERror | UVAL | NOEXists |
INSufficient
STERror: start error
SPERror: stop error
PRERror: parity error
UVAL: unexpected value
NOEXists: byte does not exist
INSufficient: the frame is not completely contained in the acqui-
sition. The decoded part of the frame is valid.
*RST: OK
Usage: Query only
BUS<m>:LIN:FRAMe<n>:IDValue?
Returns the identifier value of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
Protocols