EasyManua.ls Logo

Rohde & Schwarz RTE Series - Page 1542

Rohde & Schwarz RTE Series
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1542User Manual 1326.1032.02 ─ 20
<n> *
Selects the frame.
Return values:
<IdentifierValue> To set the value format, use FORMat:BPATtern.
The values below range, increment and reset are decimal
values.
Range: 0 to 63
*RST: 0
Usage: Query only
BUS<m>:LIN:FRAMe<n>:IDPValue?
Returns the value of the identifier parity bits of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<IdtfParityValue> To set the value format, use FORMat:BPATtern.
The values below range, increment and reset are decimal
values.
Range: 0 to 3
*RST: 0
Usage: Query only
BUS<m>:LIN:FRAMe<n>:SYSTate?
Returns the state of the sync field for the specified frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<SyncState> OK | STERror | SPERror | UVAL | NOEXists | INSufficient
STERror: start error
SPERror: stop error
UVAL: unexpected value
NOEXists: byte does not exist
INSufficient: the frame is not completely contained in the acqui-
sition. The decoded part of the frame is valid.
*RST: OK
Protocols

Table of Contents

Other manuals for Rohde & Schwarz RTE Series

Related product manuals