Remote control commands
R&S
®
RTE
1549User Manual 1326.1032.02 ─ 20
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with SEARch:TRIGger:LIN:DMIN and SEARch:
TRIGger:LIN:DMAX.
*RST: EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:LIN:DMIN <SearchName>,<DataPattern>
SEARch:TRIGger:LIN:DMIN? <SearchName>
Specifies a data pattern, or sets the the start value of a data pattern range.
Parameters:
<DataPattern> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076.
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:LIN:DMAX <SearchName>,<DataPatternTo>
SEARch:TRIGger:LIN:DMAX? <SearchName>
Sets the the end value of an identifier range if SEARch:TRIGger:LIN:DCONdition
is set to INRange or OORange.
Parameters:
<DataPatternTo> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076.
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:LIN:BORDer <SearchName>,<Endianness>
SEARch:TRIGger:LIN:BORDer? <SearchName>
Sets the byte order (endianness) of the data transfer.
Parameters:
<Endianness> BENDian | LENDian
BENDian
Big endian, data is analyzed and evaluated in the order of
reception.
LENDian
Little endian, the instrument reads the complete data, reverses
the byte order of the data, and compares it with the specified
data word.
*RST: BENDian