Remote control commands
R&S
®
RTE
1560User Manual 1326.1032.02 ─ 20
<n> 1 | 2
1 = threshold high, 2 = threshold low
Parameters:
<THResholds> Differential or absolute voltage level, depending on the source
type. See BUS<m>:FLXRay:SRCType on page 1558.
BUS<m>:FLXRay:THENable <ThresholdEnable>
Sets the threshold for the enable line if the source type is LOGic.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<ThresholdEnable> Range: -12 to 12
Increment: 0.1
*RST: 2.65
Default unit: V
BUS<m>:FLXRay:THData <ThresholdData>
Sets the threshold for the data line if the source type is LOGic.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<ThresholdData> Range: -12 to 12
Increment: 0.1
*RST: 2.35
Default unit: V
BUS<m>:FLXRay:PRSingle <PresetSingleEnd>
Selects a default threshold voltage if BUS<m>:FLXRay:SRCType is set to SINGle.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<PresetSingleEnd> MV150 | MV200 | MV250 | MV300 | MANual
MV150 | MV200 | MV250 | MV300
2.5 ± 0.15 V; 2.5 ± 0.2 V; 2.5 ± 0.25 V; 2.5 ± 0.3 V, respectively
MANual
Manual setting of user-defined values with BUS<m>:FLXRay:
THReshold<n> on page 1559.
*RST: MV150
Protocols