Remote control commands
R&S
®
RTE
1576User Manual 1326.1032.02 ─ 20
Return values:
<HdrCksumVal> To set the value format, use FORMat:BPATtern on page 1083.
The values below range, increment and default are decimal
values.
Range: 0 to 2047
*RST: 0
Usage: Query only
BUS<m>:FLXRay:FRAMe<n>:FCSTate?
Returns the state of the cyclic redundancy check code of the frame data.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<ChecksumState> OK | UVAL | INSufficient
OK: the CRC is valid.
UVAL: unexpected value
INSufficient: the frame is not completely contained in the acqui-
sition. The decoded part of the frame is valid.
*RST: OK
Usage: Query only
BUS<m>:FLXRay:FRAMe<n>:FCValue?
Returns the cyclic redundancy check code of the frame CRC.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<ChecksumValue> To set the value format, use FORMat:BPATtern on page 1083.
The values below range, increment and default are decimal
values.
Range: 0 to 16777215
*RST: 0
Usage: Query only