EasyManua.ls Logo

Rohde & Schwarz RTE Series - Page 1580

Rohde & Schwarz RTE Series
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1580User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:FLXRay:DCONdition <SearchName>,<DataOperator>
SEARch:TRIGger:FLXRay:DCONdition? <SearchName>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<DataOperator> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
OFF = ANY
The data pattern is not relevant for the search condition.
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with SEARch:TRIGger:FLXRay:DMIN.
INRange | OORange
In range, Out of range: Set the minimum and maximum value of
the range with SEARch:TRIGger:FLXRay:DMIN and SEARch:
TRIGger:FLXRay:DMAX.
*RST:
EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:DMAX <SearchName>,<DataPatternTo>
SEARch:TRIGger:FLXRay:DMAX? <SearchName>
Sets the the end value of an data range if SEARch:TRIGger:FLXRay:DCONdition
is set to INRange or OORange.
Parameters:
<DataPatternTo> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:DMIN <SearchName>,<DataPattern>
SEARch:TRIGger:FLXRay:DMIN? <SearchName>
Specifies a data pattern, or sets the the start value of a data pattern range.
Parameters:
<DataPattern> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
Protocols

Table of Contents

Other manuals for Rohde & Schwarz RTE Series

Related product manuals