Remote control commands
R&S
®
RTE
1581User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:FLXRay:DPOPerator <SearchName>,<DataPosOperator>
SEARch:TRIGger:FLXRay:DPOPerator? <SearchName>
Sets the operator for the data position. You can defined an exact position, or a position
range.
Parameters:
<DataPosOperator> ANY | OFF | EQUal | GETHan | INRange | RANGe
OFF = ANY
The data position is not relevant for the search condition.
EQUal | GETHan
Equal, Greater or equal than. These conditions require one data
position to be set with SEARch:TRIGger:FLXRay:DPOSition
INRange = RANGe
In range: Set the minimum and maximum value of the range with
SEARch:TRIGger:FLXRay:DPOSition and SEARch:
TRIGger:FLXRay:DPTO.
*RST: EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:DPOSition <SearchName>,<DataPosition>
SEARch:TRIGger:FLXRay:DPOSition? <SearchName>
Sets the number of data bytes to be skipped after start of the payload segment.
Parameters:
<DataPosition> Range: 0 to 255
Increment: 1
*RST: 0
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:DPTO <SearchName>,<DataPositionTo>
SEARch:TRIGger:FLXRay:DPTO? <SearchName>
Defines the last byte of interest, if the position operator SEARch:TRIGger:FLXRay:
DPOPerator defines a range.
Parameters:
<DataPositionTo> Range: 0 to 255
Increment: 1
*RST: 0
Parameters for setting and query:
<SearchName>
Protocols