Remote control commands
R&S
®
RTE
1635User Manual 1326.1032.02 ─ 20
BUS<m>:ARINc:PRESet <Preset>
Sets the default threshold voltage.
Suffix:
<m>
.
1..4
Parameters:
<Preset> V25 | V5 | MANual
*RST: V5
BUS<m>:ARINc:THReshold:HIGH <ThresholdHigh>
Sets the high threshold level of the signal.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdHigh> Range: 0 to 12
Increment: 0.1
*RST: 5
Default unit: V
BUS<m>:ARINc:THReshold:LOW <ThresholdLow>
Sets the low threshold level of the signal.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdLow> Range: -12 to 0
Increment: 0.1
*RST: -5
Default unit: V
18.17.11.2 Trigger
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
TRIGger<m>:SOURce[:SELect] is set to SBUS.
The sources of the serial bus are channel signals: use BUS<m>:...:SOURce
commands.
Decoding is enabled: BUS<m>[:STATe] is set to ON.
TRIGger<m>:ARINc:TYPE............................................................................................1636
TRIGger<m>:ARINc:DATA:CONDition............................................................................1636
TRIGger<m>:ARINc:DATA:MIN..................................................................................... 1636
Protocols