Remote control commands
R&S
®
RTE
1720User Manual 1326.1032.02 ─ 20
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SENT:TRANsmission <SearchName>,<TransmSequence>
SEARch:TRIGger:SENT:TRANsmission? <SearchName>
Enables the search for a transmission sequence, which is combination of status and
data conditions.
To set the transmission sequence conditions, use SEARch:TRIGger:SENT:TTYPe
and SEARch:TRIGger:SENT:STATus.
To set the data condition, use SEARch:TRIGger:SENT:TDCN, TRIGger<m>:SENT:
TDCN, SEARch:TRIGger:SENT:TDMN and SEARch:TRIGger:SENT:TDMX.
Parameters:
<TransmSequence> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SENT:SMSG <SearchName>,<SerialMessage>
SEARch:TRIGger:SENT:SMSG? <SearchName>
Enables the search in a serial message, which is combination of identifier and data
conditions.
To select the sequence condition, use SEARch:TRIGger:SENT:STYPe.
To set the identifier condition for the serial message, use SEARch:TRIGger:SENT:
SICN, SEARch:TRIGger:SENT:SIMN and SEARch:TRIGger:SENT:SIMX.
To set the data condition, use SEARch:TRIGger:SENT:SDCN, SEARch:TRIGger:
SENT:SDMN and SEARch:TRIGger:SENT:SDMX.
Parameters:
<SerialMessage> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SENT:ERRor <SearchName>,<ErrorCondition>
SEARch:TRIGger:SENT:ERRor? <SearchName>
Enables the search for specified errors.
Define the error types with SEARch:TRIGger:SENT:PULSeerror, SEARch:
TRIGger:SENT:PPERioderror, SEARch:TRIGger:SENT:FORMerror
on page 1726 and SEARch:TRIGger:SENT:CRCerror.
Protocols