EasyManua.ls Logo

Rohde & Schwarz RTE Series - Configuration; MDIO (Option R&S RTE-K55)

Rohde & Schwarz RTE Series
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1776User Manual 1326.1032.02 ─ 20
18.17.16 MDIO (option R&S RTE-K55)
Configuration.......................................................................................................1776
Trigger.................................................................................................................1778
Decode results.................................................................................................... 1781
Search settings................................................................................................... 1784
Search results..................................................................................................... 1787
18.17.16.1 Configuration
In all BUS<m>:MDIO commands, the suffix <m> selects the serial bus.
BUS<m>:MDIO:CLOCk:SOURce...................................................................................1776
BUS<m>:MDIO:DATA:SOURce.....................................................................................1776
BUS<m>:MDIO:CLOCk:THReshold:HIGH...................................................................... 1777
BUS<m>:MDIO:CLOCk:THReshold:LOW.......................................................................1777
BUS<m>:MDIO:DATA:THReshold:HIGH.........................................................................1777
BUS<m>:MDIO:DATA:THReshold:LOW......................................................................... 1778
BUS<m>:MDIO:PRESet............................................................................................... 1778
BUS<m>:MDIO:COUPling............................................................................................ 1778
BUS<m>:MDIO:CLOCk:SOURce <SourceClock>
Selects the source for the clock line (management data clock, MDC). Permitted selec-
tions are the analog channels "C1""C4" and the digital channels "D0""D15".
Suffix:
<m>
.
1..4
Parameters:
<SourceClock> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
data and clock lines. For triggering on a serial bus, analog or
digital input channels are required.
See Chapter 18.4.2, "Waveform parameter", on page 1074
*RST: C2W1
BUS<m>:MDIO:DATA:SOURce <SourceData>
Selects the source for the data signal. Permitted selections are the analog channels
"C1""C4" and the digital channels "D0""D15", but not the same as for "Clock".
Suffix:
<m>
.
1..4
Protocols