Remote control commands
R&S
®
RTE
1780User Manual 1326.1032.02 ─ 20
TRIGger<m>:MDIO:FRAMetype <FrameType>
Selects the Type of Frame code (or OP code, OpCode, operation code); available only
in trigger type "Data".
Note that the same OpCode may have different meanings in Clause 22 and Clause 45.
Parameters:
<FrameType> OP00 | OP01 | OP10 | OP11 | OPXX
OP00
Address frame (in Clause 45, only)
OP01
Write frame (in Clause 22 or Clause 45)
OP10
Read frame (in Clause 22) or Post-Read increment address
frame (in Clause 45)
OP11
Read frame (in Clause 45)
OPXX
Any frame type
*RST:
OPXX
Note that the user interface shows interpretations of the numeri-
cal OpCode values corresponding to Clause 45. Clause 22 is
not represented by this interpretation.
TRIGger<m>:MDIO:PHYS <PhyAddr>
Sets the physical address (in Clause 22) or port address (in Clause 45) of the frame
pattern (5 bits); available only in trigger type "Data".
Parameters:
<PhyAddr>
TRIGger<m>:MDIO:REGI <RegAddr>
Sets the register address (in Clause 22) or device address (in Clause 45) of the frame
pattern (5 bits); available only in trigger type "Data".
Parameters:
<RegAddr>
TRIGger<m>:MDIO:DATA <Data>
Defines the 16-bit payload data pattern (both in Clause 22 or Clause 45) or the
address pattern (in Clause 45, only) to trigger for; available only in trigger type "Data".
Parameters:
<Data>
Protocols