Remote control commands
R&S
®
RTE
1783User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..4
<n> *
Return values:
<Start> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:MDIO:WORD<n>:STATe?
Returns the state of the frame in the present acquisition of the selected serial bus.
Suffix:
<m>
.
1..4
<n> *
Return values:
<State> OK | UNSYN | UNSYncronized | OPCO | TA_ERROR |
INComplete | SHORT | SHORt
OK
No error detected
UNSYN = UNSYncronized
UNSYncronized happens when bits are found, but they are not
correlated with any synchronization sequence. We don't know
what these bits are, but they are there. These bits receive a flag,
but they are not decoded.
OPCO
OPcode Error
TA_ERROR
Tirnaround time error
INComplete
Incomplete Frame
SHORT = SHORt
Length Error
*RST:
OK
Usage: Query only
BUS<m>:MDIO:WORD<n>:STOP?
Returns the stop time of the frame from the selected word within the search result.
Suffix:
<m>
.
1..4
<n> *