Remote control commands
R&S
®
RTE
1823User Manual 1326.1032.02 ─ 20
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one ET value to
be set using SEARch:TRIGger:USB:TMIN.
INRange | OORange
In range, out of range. These conditions require a range of ET
values to be set using SEARch:TRIGger:USB:TMIN and
SEARch:TRIGger:USB:TMAX.
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:TMIN
<SearchName>,<ET>
SEARch:TRIGger:USB:TMIN? <SearchName>
Specifies an endpoint type (ET), or sets the start value of an ET range.
Parameters:
<ET>
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:TMAX <SearchName>,<ETTo>
SEARch:TRIGger:USB:TMAX? <SearchName>
Sets the the end value of an endpoint type (ET) range if TRIGger<m>:USB:
TCONdition is set to INRange or OORange.
Parameters:
<ETTo>
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:TOKen <SearchName>,<USBTokenType>
SEARch:TRIGger:USB:TOKen? <SearchName>
Defines, which token packet type is searched for: "OUT", "IN", "SOF", "SETUP", or
"Any" token packet.
Parameters:
<USBTokenType> ANY | OUT | IN | SOF | SETup
ANY
Searches for any of the token packet types listed below.
OUT
Searches for an OUT token.
IN
Searches for an IN token.