EasyManua.ls Logo

Rohde & Schwarz RTE Series - Channels; Vertical Settings

Rohde & Schwarz RTE Series
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Acquisition and waveform setup
R&S
®
RTE
185User Manual 1326.1032.02 ─ 20
To acquire the maximum number, enable "Acquire maximum". The maximum num-
ber of acquisitions is shown in the "Required" field.
Acquire a given number of acquisitions.
Enter the number in the "Required" field.
The acquisition count (Average count (N-single count)) is always set to the required
number of acquisitions. Thus you can acquire exactly one fast segmentation acquisi-
tion series with [RUN N× SINGLE]. The [RUN CONT] key works in the same way as
[RUN N× SINGLE], it stops acquisition when the series is completed.
You can stop the running acquisition before the series is completed.
The number of acquired waveforms is shown in "Available" and can be displayed with
"Show history".
Remote command:
ACQuire:SEGMented:MAX on page 1116
Replay time
Defines the display speed of the fast segmentation acquisition series. Display starts
after the series has been captured completely.
See also: "Replay time per acq." on page 314.
Show history
Enables the history mode and displays the history viewing functions in the "Fast Seg-
mentation" tab.
See also: Chapter 7.4.2.1, "Viewer", on page 313.
5.3 Vertical settings
The "Vertical" menu contains all channel-dependent settings and information.
Channels...............................................................................................................185
Coupled channels................................................................................................. 188
Power calculation..................................................................................................189
5.3.1 Channels
Access: "Vertical" menu > "Channels"
The "Channels" tab provides all basic vertical settings. The channels are listed in verti-
cal tabs at the left side of the dialog box.
Vertical settings

Table of Contents

Other manuals for Rohde & Schwarz RTE Series

Related product manuals