Remote control commands
R&S
®
RTE
1860User Manual 1326.1032.02 ─ 20
*RST: EQUal
TRIGger<m>:SWIRe:DATA:MAX <DataPatternMax>
Sets the the end value of a data for the data pattern range if TRIGger<m>:SWIRe:
DATA:CONDition is set to INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<DataPatternMax>
TRIGger<m>:SWIRe:DATA:MIN
<DataPatternMin>
Specifies a data for the data pattern, or sets the the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<DataPatternMin>
TRIGger<m>:SWIRe:ERRor:ESC <ErrorESC>
Enables searching for an escape error.
Suffix:
<m>
.
1..3
Parameters:
<ErrorESC> ON | OFF
*RST: ON
TRIGger<m>:SWIRe:ERRor:PARity <ErrorParity>
Enables triggering on a parity error (even parity).
Suffix:
<m>
.
1..3
Parameters:
<ErrorParity> ON | OFF
*RST: ON
TRIGger<m>:SWIRe:TIME:CONDition <TimeOperator>
Set the condition for the data value for the time code. You can define an exact value or
a value range.
Protocols