Remote control commands
R&S
®
RTE
1917User Manual 1326.1032.02 ─ 20
Parameters:
<SetupTime> Range: -99.8E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
TRIGger<m>:PARallel:DATatoclock:HTIMe <HoldTime>
Sets the minimum time after the clock edge while data should be stable and not
change its state.
See also: "Hold time" on page 915
Parameters:
<HoldTime> Range: -99.8E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
18.18.4.6 State trigger
See also:
●
TRIGger<m>:PARallel:STATe:CSOurce:VALue on page 1913
●
TRIGger<m>:PARallel:STATe:EXPRession[:DEFine] on page 1913
TRIGger<m>:PARallel:STATe:CSOurce:EDGE................................................................1917
TRIGger<m>:PARallel:STATe:BIT<0..15>....................................................................... 1917
TRIGger<m>:PARallel:STATe:CSOurce:EDGE <Slope>
Sets the edge of the clock signal. The crossing of the clock edge and the logical
threshold defines the time at which the logical states and the bus value are analyzed.
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
TRIGger<m>:PARallel:STATe:BIT<0..15> <Bit>
Sets the required state for each digital channel that is used in the bus.
Parameters:
<Bit> HIGH | LOW | DONTCARE | DONTcare
Bit value: 1 (HIGH), 0 (LOW), or X (DONTCARE = DONTcare)
Usage: Asynchronous command
Mixed signal option (MSO, R&S RTE-B1)