Remote control commands
R&S
®
RTE
1919User Manual 1326.1032.02 ─ 20
Usage: Asynchronous command
TRIGger<m>:PARallel:PATTern:TIMeout[:TIME] <Time>
Defines the time limit for the timeout at which the instrument triggers.
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 100E-9
Default unit: s
Usage: Asynchronous command
TRIGger<m>:PARallel:PATTern:WIDTh:RANGe <WidthRangeMode>
Selects how the range of a pulse width is defined if TRIGger<m>:PARallel:
PATTern:MODE is set to WIDTh.
Parameters:
<WidthRangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers when the pattern comes false inside a given time
range. The time limit is defined by TRIGger<m>:PARallel:
PATTern:WIDTh[:WIDTh] and TRIGger<m>:PARallel:
PATTern:WIDTh:DELTa (Width
±
Delta).
OUTSide
Triggers when the pattern comes false before or after the given
time range. The time limit definition is the same as for WITHin
range.
SHORter | LONGer
Triggers when the pattern comes false before or after the given
width has expired. Width is set with TRIGger<m>:PARallel:
PATTern:WIDTh[:WIDTh].
*RST:
WITHin
Usage: Asynchronous command
TRIGger<m>:PARallel:PATTern:WIDTh[:WIDTh] <Width>
The effect depend on the setting of the TRIGger<m>:PARallel:PATTern:WIDTh:
RANGecommand.
For the ranges SHOrter and LONGer, the width defines the maximum and minimum
time limit, respectively.
For the ranges WITHin and OUTSide, the width defines the center of a range which is
defined by the limits "±Delta".
Mixed signal option (MSO, R&S
RTE-B1)