Parameters:
<CarrierFreq> Range: 1E-3 to 100E+6
Increment: 10
*RST: 1E+6
Default unit: Hz
WGENerator<m>:MODulation:CARRier:PERiod <CarrierPeriod>
Sets the period of the carrier signal for a modulation waveform.
Suffix:
<m>
.
1..2
Parameters:
<CarrierPeriod> Range: 8E-9 to 1000
Increment: 10
*RST: 1E-6
Default unit: s
WGENerator<m>:MODulation:AM[:FUNCtion] <SignalType>
Sets the type of the modulation signal for the amplitude modulation, if
WGENerator<m>:MODulation:TYPE is set to AM.
Suffix:
<m>
.
1..2
Parameters:
<SignalType> SINusoid | SQUare | RAMP
*RST: SINusoid
WGENerator<m>:MODulation:FM:DCYCle <SquareDutyCycle>
Sets the duty cycle of a square waveform for frequency modulation, if
WGENerator<m>:MODulation:TYPE is set to FM.
Suffix:
<m>
.
1..2
Parameters:
<SquareDutyCycle> Range: 10 to 90
Increment: 1
*RST: 50
Default unit: %
WGENerator<m>:MODulation:FM:DEViation <Deviation>
Sets the frequency modulation deviation.