Acquisition and waveform setup
R&S
®
RTE
212User Manual 1326.1032.02 ─ 20
Current probes R&S RT-ZCxx
The current probes R&S RT-ZCxx have BNC connectors. They are known to the
R&S RTE as predefined probes, see Chapter 5.5.6, "Setup for predefined probes",
on page 209. Demagnetizing and zero adjustment is done on the probe, see the pro-
be's User Manual for details. Make sure to demagnetize and adjust the probe before
taking measurements.
Figure 5-16: Probe setup for current probes R&S
RT-ZC10
Current probes R&S RT-ZCxxB
Current probes R&S RT-ZCxxB have a Rohde & Schwarz probe interface; they are
powered and remotely controlled by the oscilloscope.
When the probe is connected, demagnetization is performed automatically.