Acquisition and waveform setup
R&S
®
RTE
227User Manual 1326.1032.02 ─ 20
For using internal shunts, the circuit under test needs to be interrupted so that the cur-
rent can flow through the probe.
Remote command:
ZVC:Z<m>:I<n>:SHUNt:MAXCurrent on page 1148
External shunt mode
If Shunt mode is set to "External shunt", include the settings for the external shunt
mode.
Max voltage ← External shunt mode
Selects the maximum voltage at the external shunt.
Remote command:
ZVC:Z<m>:I<n>:SHUNt:MAXVoltage on page 1149
Shunt value ← External shunt mode
Sets the shunt value of the external shunt resistor.
Remote command:
ZVC:Z<m>:I<n>:SHUNt:EVALue on page 1148
Max current ← External shunt mode
Displays the maximum current.
Remote command:
ZVC:Z<m>:I<n>:SHUNt:MXCValue? on page 1149
Reset Overload ← External shunt mode
Resets the overload indication at the probe.
Remote command:
ZVC:Z<m>:I<n>:OVERload:RSTO on page 1146
5.7 Differential signals
With R&S RTE, you can easily analyze differential signals using single-ended probes,
or even cables. The instrument processes the input of single-ended probes on analog
channels and creates the differential and common mode waveforms. Similar to
Rohde & Schwarz modular probes, you can display the waveforms of differential volt-
age, common mode voltage, positive single-ended voltage and negative single-ended
voltage. Triggering on differential signals is also possible.
Differential signals