Acquisition and waveform setup
R&S
®
RTE
229User Manual 1326.1032.02 ─ 20
"Differential"
Displays the differential signal as a new waveform.
"Common"
Displays the common mode signal as a new waveform.
Remote command:
DIFFerential<m>:AOUTput on page 1158
DIFFerential<m>:BOUTput on page 1158
Vertical scale, Offset
Vertical scale and offset of differential and common mode waveforms can be set
directly in the differential setup, or in the "Vertical Setup" of differential signals. See:
"Vertical scale" on page 229 and "Offset" on page 230.
Vertical settings of P and N output are the vertical settings of the input channels.
5.7.2 Vertical setup of differential signals
Access: "Vertical" menu > "Differential Signals" > "Vertical Setup"
Vertical scale
Sets the vertical scale of differential and common mode waveforms.
Vertical settings of P and N output are the vertical settings of the input channels.
Remote command:
DIFFerential<m>:COMMon:SCALe on page 1158
DIFFerential<m>:DIFFerential:SCALe on page 1158
Scale coupling
If enabled, the vertical scales of P, N, differential and common mode waveforms are
coupled. The scales are related as follows:
VertScale
P
= VertScale
N
= VertScale
CM
= VertScale
Diff
/2
You can disable the scale coupling and set an individual scale for each waveform.
Differential signals