Acquisition and waveform setup
R&S
®
RTE
237User Manual 1326.1032.02 ─ 20
5.10.7 Using fast segmentation
Fast Segmentation reduces the dead time between two waveform acquisition cycles.
The settings are described in Chapter 5.2.3, "Fast segmentation", on page 183.
1. On the "Horizontal" menu, tap "Fast Segmentation".
2. Tap "Enable" to activate the Fast Segmentation mode.
3. If you want to sample the maximum number of acquisitions in a series, select
"Acquire maximum".
If you want to capture a defined number of acquisitions, disable "Acquire maxi-
mum" and enter the "Required" number of acquisitions.
4. Set the "Replay time", the display time of each acquisition.
5.10.8 Using digital filters
Before using digital filters, you determine if you want to filter input channels only or if
the trigger signal is filtered too. The filter settings depend on this decision.
For details on filter settings and dependencies, see Chapter 5.8, "Digital filter setup",
on page 230.
1. On the "Vertical" menu, tap "Digital Filter Setup".
2. To filter the input channels, enable "Acquisition RF reject".
3. To filter the input channels, enable "Trigger RF reject".
4. Set the frequency limit for the filter: "Bandwidth".
This limit is applied to the trigger channel and to the acquisition channels enabled
for filtering.
Setting up the waveform