Triggers
R&S
®
RTE
239User Manual 1326.1032.02 ─ 20
Trigger sequence
A complex trigger sequence joins two or more separate trigger conditions with an
optional delay time and an optional reset time or reset condition. Similar setups are
also known as multi-step trigger or A/B trigger.
Trigger information
Information on the most important trigger settings is shown in the trigger label on top of
the signal bar. If you double-tap the trigger label, the "Trigger" dialog box opens. The
label shows:
Trigger mode
Trigger type, edge/polarity and trigger source, for A- and B-trigger
Trigger level
If the trigger mode is normal, and no trigger has been found for longer than one sec-
ond, the trigger state is shown in the signal bar. The state indicates the waiting for trig-
ger time. For long timebases, it shows the remaining pretrigger time, and after the trig-
ger the time until the acquisition is completed.
6.2 Setting up the trigger
This chapter provides step-by-step procedures for the important stages of trigger
setup.
6.2.1 Configuring a simple trigger
Prerequisites:
Horizontal and vertical settings are set appropriately to the signals.
The acquisition is running, the [RUN CONT] key lights green.
For details on settings, see Chapter 6.3, "Trigger types", on page 241.
Setting up the trigger