Triggers
R&S
®
RTE
255User Manual 1326.1032.02 ─ 20
Data level
Sets the voltage level for the data signal. At this level, the setup and hold time is mea-
sured.
Remote command:
TRIGger<m>:LEVel<n>[:VALue] on page 1166
Couple levels (Trigger level and hysteresis coupling)
Sets the trigger levels and hysteresis values for all channels to the values of the cur-
rently selected trigger source. The function affects only the levels defined for the
selected condition. The hysteresis of the external trigger input is an independent value,
and it is not affected by level coupling.
Remote command:
TRIGger<m>:SCOupling on page 1184
Setup time
Sets the minimum time before the clock edge while the data signal must stay steady
above or below the data level.
The setup time can be negative. In this case, the hold time is always positive. If you set
a negative setup time, the hold time is adjusted by the instrument.
Remote command:
TRIGger<m>:DATatoclock:STIMe on page 1185
Hold time
Sets the minimum time after the clock edge while the data signal must stay steady
above or below the data level.
The hold time can be negative. In this case, the setup time is always positive. If you set
a negative hold time, the setup time is adjusted by the instrument.
Remote command:
TRIGger<m>:DATatoclock:HTIMe on page 1184
6.3.11 State
The state trigger is a qualified edge trigger. It combines the edge trigger settings with
trigger qualification.