Triggers
R&S
®
RTE
266User Manual 1326.1032.02 ─ 20
Hysteresis mode
Selects how the hysteresis is set.
"Auto"
This is the recommended mode. The hysteresis is set by the instru-
ment to reject the internal noise of the instrument.
"Manual"
The hysteresis is defined directly in absolute or relative values.
Remote command:
TRIGger<m>:LEVel<n>:NOISe[:STATe] on page 1201
Scale mode
Selects whether the hysteresis is defined in absolute or relative values. The setting is
available only in manual hysteresis mode.
Remote command:
TRIGger<m>:LEVel<n>:NOISe:MODE on page 1201
Absolute hysteresis
Defines a range in absolute values around the trigger level. If the signal jitters inside
this range and crosses the trigger level thereby, no trigger event occurs.
Remote command:
TRIGger<m>:LEVel<n>:NOISe:ABSolute on page 1202
Relative hysteresis
Defines a range in divisions around the trigger level. If the signal oscillates inside this
range and crosses the trigger level thereby, no trigger event occurs.
Noise Reject