Measurements
R&S
®
RTE
348User Manual 1326.1032.02 ─ 20
Remote command:
Source is defined by suffix <m> in "REFLevel" subsystem, see Chapter 18.12.13, "Ref-
erence levels", on page 1308
Reference level mode
Defines whether the reference level is configured manually or automatically.
Remote command:
REFLevel<m>:LDETection on page 1309
Level definition
In manual reference level mode, the setting defines whether the reference is config-
ured using absolute or relative values.
Remote command:
REFLevel<m>:LMODe on page 1309
Relative levels
Sets the lower, middle and upper reference levels, defined as percentages of the signal
amplitude.
Available relative levels:
5/50/95
10/50/90
20/50/80
User defined: Enter "Upper ref level", "Middle ref level", and "Lower ref level".
For example, for "5/50/95" the levels are set to the following values:
Lower reference level = 5% of the signal amplitude
Middle reference level = 50% of the signal amplitude
Upper reference level = 95% of the signal amplitude
Remote command:
REFLevel<m>:RELative:MODE on page 1309
Upper ref level, Middle ref level, Lower ref level
Define the reference levels in percent, if "Relative levels" is set to "User-defined".
Remote command:
REFLevel<m>:RELative:UPPer on page 1313
REFLevel<m>:RELative:MIDDle on page 1313
REFLevel<m>:RELative:LOWer on page 1313
High
Sets the high signal level.
The high signal level is set in manual reference level mode for absolute level definition.
Remote command:
REFLevel<m>:ABSolute:HIGH on page 1311
Low
Sets the low signal level.
The low signal level is set in manual reference level mode for absolute level definition.