Measurements
R&S
®
RTE
391User Manual 1326.1032.02 ─ 20
Scale mode
Defines when the points of a long-term measurement are created.
If statistics are enabled, each long-term measurement point shows the statistical mean
and standard deviation of the results measured during the defined period.
If statistics are disabled, the first measurement result of each period is taken as long-
term measurement point.
"Time"
Sets one long-term measurement point for the time defined in "Time/
point".
"Waveforms"
Sets one long-term measurement point for several acquired wave-
forms. The number is defined in "Waveforms/point".
"Measure-
ments"
Sets one long-term measurement point for several measurement
results. The number is defined in "Measurements/point".
Remote command:
MEASurement<m>:STATistics:MODE on page 1291
Time / point
Defines the time to create one point of the long-term measurement. The "Time / point"
value is a lower time limit. The actual time between two points depends on the acquisi-
tion and postprocessing time.
The long-term measurement is not a data logger with equidistant points as the time
between two points varies.
This setting is only available if "Scale mode" is set to "Time".
Remote command:
MEASurement<m>:STATistics:RTIMe on page 1292
Measurement time
Defines the total duration of the long-term measurement: Time/point * Number of
points.
This setting is only available if "Scale mode" is set to "Time".
Remote command:
MEASurement<m>:LTMeas:TIME on page 1295
Wfms / point
Defines the number of measured waveforms from which one point of the long-term
measurement is created.
This setting is only available if "Scale mode" is set to "Waveforms".
Remote command:
MEASurement<m>:STATistics:RCOunt on page 1291
Measurements / point
Defines the number of measurement results from which one point of the long-term
measurement is created.
This setting is only available if "Scale mode" is set to "Measurements".
Remote command:
MEASurement<m>:STATistics:RMEascount on page 1292