Mask testing
R&S
®
RTE
434User Manual 1326.1032.02 ─ 20
Remote command:
MTESt:ONViolation:STOP on page 1344
Print
Prints a screenshot including the mask test results to the printer defined in the "Print"
dialog box (see Printer).
Remote command:
MTESt:ONViolation:PRINt on page 1344
Save Wfm
Saves the failed waveform as a reference waveform to the file specified in [FILE] >
"Save/Recall" > "Waveform".
Remote command:
MTESt:ONViolation:SAVewaveform on page 1344
Trigger Out Pulse
Creates a pulse on the [EXT TRIGGER OUT] connector on mask violation or success-
ful completion of the test cycle. The minimum time difference between two trigger out
pulses is 30 ms because the instrument detects mask violation at display update.
Events with a higher frequency are not captured completely.
If this event is enabled and the mask test is running, the trigger control option "Enable
trigger out" is disabled. Thus, the trigger out pulse is provided only on mask test result
but not when a trigger occurs. The pulse is provided always with the minimum delay of
800 ns, the "Delay" cannot be set.
Remote command:
MTESt:ONViolation:TRIGgerout on page 1345
Report
Creates and saves a report using the settings defined in "File" menu > "Report Setup".
Remote command:
MTESt:ONViolation:REPort on page 1345
Start Executable
Starts an external application. Tap "Config Executable" to set the application path and
parameters.
See: Chapter 4.5, "External application", on page 157.
Remote command:
MTESt:ONViolation:RUNexec on page 1345
Reset
Clears all totals and results in all "Mask Test" result boxes.
Remote command:
MTESt:RST on page 1333
Mask test settings