Mask testing
R&S
®
RTE
435User Manual 1326.1032.02 ─ 20
10.2.4 Mask display
Access: [MASKS] > "Mask Display"
The "Mask Display" tab contains all settings for mask and hit display.
Show mask
Switches the display of all mask segments on or off.
Show labels
Switches the display of the mask test name on or off.
To change the name of the mask test, open the "Test Definition" tab, double-tap the
mask test subtab and enter the new mask test name.
Remote command:
MTESt:LABel on page 1346
MTESt:REName on page 1346
Waveform style
See: "Style" on page 145.
Highlight hits
If selected, the mask hits are highlighted on the screen. You can define the color and
the time of the hit display.
Remote command:
MTESt:HIGHlight:STATe on page 1347
Infinite higlight
If selected, the mask hits are shown for an unlimited period of time.
Remote command:
MTESt:HIGHlight:INFinite on page 1346
Highlight time
Sets the time for which the mask hits are shown.
Remote command:
MTESt:HIGHlight:TIME on page 1347