Protocol analysis
R&S
®
RTE
602User Manual 1326.1032.02 ─ 20
See also: Chapter 13.1.1, "Decode - general settings", on page 514.
LIN standard
Selects the version of the LIN standard that is used in the DUT. The setting mainly
defines the checksum version used during decoding.
The most common version is LIN 2.x. For mixed networks, or if the standard is
unknown, set the LIN standard to "Auto".
Remote command:
BUS<m>:LIN:STANdard on page 1531
Data
Sets the source waveform of the data line.
Usually, the source is one of the analog channels. Reference and math waveforms are
available as source if the trigger source is one of the analog channels but not the serial
bus.
Alternatively, digital channels can be used if MSO option R&S RTE-B1 is installed. Dig-
ital and analog channels cannot be used at the same time.
For triggering on a serial bus, analog or digital channel sources are required.
Remote command:
BUS<m>:LIN:DATA:SOURce on page 1529
Threshold
Sets the threshold value for digitization of the signal. If the signal value on the line is
higher than the threshold, the signal state is high. Otherwise, the signal state is consid-
ered low if the signal value is below the threshold. The interpretation of HIGH and LOW
is defined by the Polarity.
There are several ways to set the threshold:
●
"Threshold"
Enter the value directly in the field.
●
"Full autoset"
Starts software algorithms for determining the signal threshold levels and bitrate.
LIN (option R&S
RTE-K3)