Protocol analysis
R&S
®
RTE
605User Manual 1326.1032.02 ─ 20
Trigger type
Selects the trigger type for LIN analysis.
"Start of frame
(Sync)"
Triggers on the stop bit of the sync field.
"Identifier"
Sets the trigger to one specific identifier or an identifier range. Enter
only the 6-bit identifier without parity bits, not the protected identifier.
Description of trigger type specific settings: "Identifier setup: Condi-
tion, Frame ID min, Frame ID max" on page 605.
"Identifier OR"
Sets the trigger to a combination of up to four identifiers.
Description of trigger type specific settings: "Identifier OR setup: Mon-
itor, Frame ID" on page 606
"Identifier +
Data"
Sets the trigger to a combination of identifier and data condition. The
instrument triggers at the end of the last byte of the specified data
pattern.
The identifier conditions are the same as for the "Identifier" trigger
type, see Identifier setup: Condition, Frame ID min, Frame ID max.
Data conditions are set with Data setup: Data length, Transfer, Condi-
tion, Data min, Data max.
"Wakeup
frame"
Triggers after a wakeup frame.
"Error condi-
tion"
Identifies various errors in the frame, see "Error conditions"
on page 607.
Remote command:
TRIGger<m>:LIN:TYPE on page 1532
Identifier setup: Condition, Frame ID min, Frame ID max
The identifier setup consists of the condition and one or two identifier pattern.
"Condition"
Defines the operator to set a specific identifier ("Equal" or "Not
equal") or an identifier range.
"Frame ID
min / Frame
ID"
Defines the bit pattern of the identifier. Enter only the 6-bit identifier
without parity bits, not the protected identifier.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.5, "Bit pattern editor", on page 521.
"Frame ID
max"
The second identifier pattern is required to specify a range with condi-
tions "In range" and "Out of range".
LIN (option R&S RTE-K3)