Protocol analysis
R&S
®
RTE
669User Manual 1326.1032.02 ─ 20
Synchronization Error ← Error Condition
Triggers if a sync impulse does not fulfill the technical requirements or when the trans-
mission is not valid.
Remote command:
TRIGger<m>:MILStd:ERRor:SYNC on page 1616
Manchester Coding Error ← Error Condition
Triggers if there is an error in the Manchester coding of the signal.
Remote command:
TRIGger<m>:MILStd:ERRor:MANChester on page 1616
Parity Error ← Error Condition
Checks the parity of every word and triggers if the parity is even.
Remote command:
TRIGger<m>:MILStd:ERRor:PARity on page 1616
Min Gap ← Error Condition
Checks and sets a value for the intermessage gap between the last bit of a message
and the following command word sync. The time is measured between the mid bit zero
crossings. According to the standard, the minimum idle time is 4 μs.
The minimum gap time is relevant for protocol configuration and error trigger.
If "Min Gap" is enabled, the instrument detects the specified gap during decoding. If
the trigger type "Error condition" is selected in addition, the instrument triggers when
the gap is shorter then specified.
Remote command:
BUS<m>:MILStd:MINGap:BITS on page 1608
BUS<m>:MILStd:MINGap:SELect on page 1607
TRIGger<m>:MILStd:MINGap:BITS on page 1617
TRIGger<m>:MILStd:MINGap:SELect on page 1617
MIL-1553 (option R&S
RTE-K6)