Getting started
R&S
®
RTE
75User Manual 1326.1032.02 ─ 20
4. If necessary, compensate the passive probe as described in Chapter 5.10.1,
"Adjusting passive probes", on page 234.
5. In the [TRIGGER] area of the front panel, press the [SOURCE] key. Press the key
again to switch the trigger source to "C2".
An unstable waveform is displayed. In auto mode, the instrument triggers repeat-
edly after a time interval if no real trigger occurs.
6. In the [TRIGGER] area, press the [MODE] key.
7. Check the "Trigger" settings in the upper right corner of the screen.
The trigger mode has changed to "Normal". The waveform is no longer refreshed,
and "Wait" is displayed in the trigger settings. The instrument cannot find a real
trigger event because there is no signal on channel 2.
8. Tap the "Undo" icon on the toolbar repeatedly until the trigger mode is "Auto" and
the trigger source is "CH1".
9. Press the [SLOPE] key to toggle the trigger slope.
Watch the waveform and the "Trigger" settings.
Trying out the instrument