Mixed signal option (MSO, R&S RTE-B1)
R&S
®
RTE
912User Manual 1326.1032.02 ─ 20
14.3.1.3 Width
The width trigger detects positive and/or negative pulses of a pulse width (duration)
inside or outside of a defined time limit. It can trigger on a single digital channel or a
logical combination of digital channels.
The instrument triggers at the end of the detected pulse.
Figure 14-2: Width trigger settings for trigger source = logical combination of digital channels
Polarity
Sets the polarity of a pulse to "Positive", "Negative", or "Both".
When triggering on a positive pulse, the trigger event occurs on the high to low transi-
tion of the pulse if the timing condition is true. When triggering on a negative pulse, the
trigger event occurs on the low to high transition of the pulse if the timing condition is
true.
Remote command:
TRIGger<m>:PARallel:WIDTh:POLarity on page 1914
Logical expression
Defines a logical combination of several digital channels as trigger condition if "Logic"
is set for "Source". As long as the digital signals match the logical expression (true),
the pulse is positive. Otherwise, the pulse is negative.
Remote command:
TRIGger<m>:PARallel:WIDTh:EXPRession[:DEFine] on page 1913
Range
Selects how the range of a pulse width is defined:
"Within"
Triggers on pulses inside a given time range. The time limit is defined
by Width
±
Delta.
"Outside"
Triggers on pulses shorter or longer than a given time range. The
time limit definition is the same as for "Within" range.
"Shorter"
Triggers on pulses shorter than the given "Width".
"Longer"
Triggers on pulses longer than the given "Width".
Remote command:
TRIGger<m>:PARallel:WIDTh:RANGe on page 1914
Trigger