8. In the results table, tap the row of the search result that you want to display in the
search zoom diagram.
3.3.10 Performing a mask test
In the following example, you perform a mask test to determine whether the signal
exceeds a rectangular area.
1. To restore the default signal channel settings, press [PRESET] and [AUTOSET].
2. Tap the "Masks" icon on the toolbar.
3. Tap the corner points of the first mask segment on the touchscreen. Select the cor-
ner points of one complete pulse, with a minor offset to the inside. To finish the
mask definition, double-tap the last point.
To create an exact rectangle, draw the diagonal of the rectangle on the screen.
Trying out the instrument