Power analysis (option R&S RTE-K31)
R&S
®
RTE
999User Manual 1326.1032.02 ─ 20
Figure 16-3: Settings turn off time
Input Type
To access this setting, select the "Channels" tab.
Selects the AC or DC input type.
Remote command:
POWer:ONOFf:INPut on page 1966
Measurement Type
Selects the"Turn on" or the " Turn off " measurement.
Remote command:
POWer:ONOFf:TYPE on page 1967
Turn on
Enables the configuration of the turn on time measurement.
Steady state level-Turn on ← Turn on
Sets the percentage of the steady state level of the DC output that has to be reached.
Remote command:
POWer:ONOFf:DSON on page 1966
Trigger level on ← Turn on
Triggers the beginning of the measurements at the moment the AC or DC input voltage
reaches the set value.
Remote command:
POWer:ONOFf:ATON on page 1966
POWer:ONOFf:DTON on page 1966
Turn off
Enables the configuration of the turn off time measurement.
Power path