Trying Out the Instrument
R&S
®
RTE
41Getting Started 1326.1026.02 ─ 08
If you use a passive probe, you can filter the DC component by using the AC cou-
pling function. Then you quickly find the new trigger level, and try out the scaling
functions:
1.
Press the [Ch1] key on the front panel (in the [VERTICAL] area) to display the
"Vertical" dialog box.
2. Change the "Coupling" to "AC". Close the dialog box.
The DC component of the signal is eliminated; the waveform position moves
down vertically and is now centered on 0 V.
Changing the Waveform Scaling and Position