Trying Out the Instrument
R&S
®
RTE
65Getting Started 1326.1026.02 ─ 08
7. Select the "Test Definition" tab.
8.
Select channel 1 as the "Source".
9. Define the number of tolerable sample hits in the "Tolerance" field.
A test has failed if the number of sample hits exceeds the limit of "Violation
tolerance" hits.
10. Select the "Event Action / Reset" tab.
11. For the "Stop acquisition" action, select On violation. If the violation tolerance
is exceeded, acquisition stops.
The results of the mask test are shown in the "MaskTest" results box. Mask
hits are also indicated as red points in the mask segment in the diagram.
Performing a Mask Test