Remote control commands
R&S
®
RTE
1184User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockEdge> POSitive | NEGative | EITHer
See Chapter 18.4.3, "Slope parameter", on page 1075.
*RST: POSitive
Usage: Asynchronous command
TRIGger<m>:DATatoclock:CSOurce:LEVel
<ClockLevel>
Sets the voltage level for the clock signal. Both this command and TRIGger<m>:
DATatoclock:CSOurce:EDGE define the starting point for calculation of the setup
and hold time.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockLevel> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
TRIGger<m>:SCOupling <TrigLevSrcCpl>
Sets the trigger levels of all used channels to the level of the data line ("Data2Clock"
and serial pattern trigger), clock line (state trigger) or source line (pattern trigger).
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<TrigLevSrcCpl> ON | OFF
*RST: OFF
Usage: Asynchronous command
TRIGger<m>:DATatoclock:HTIMe <HoldTime>
Sets the minimum time after the clock edge while the data signal must stay steady
above or below the data level.
Trigger