Remote control commands
R&S
®
RTE
1203User Manual 1326.1032.02 ─ 20
<n> 1..31
Indicates the trigger source: see TRIGger<m>:LEVel<n>:
NOISe[:STATe] on page 1201.
Parameters:
<HystInDivs> Hysteresis size in divisions
Range: 0 to 5
Increment: 0.01
*RST: 0
Default unit: div
Usage: Asynchronous command
TRIGger<m>:LEVel<n>:NOISe:RELative
<HystRel>
Defines a range in percent around the trigger level. If the signal oscillates inside this
range and crosses the trigger level thereby, no trigger event occurs.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
<n> 1..31
Indicates the trigger source: see TRIGger<m>:LEVel<n>:
NOISe[:STATe] on page 1201.
Parameters:
<HystRel> Hysteresis in %. 10% = 1 div
Range: 0 to 50
Increment: 1
*RST: 0
Default unit: %
Usage: Asynchronous command
TRIGger<m>:ANEDge:NREJect <NoiseReject>
Enables the noise reject for the external trigger input.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<NoiseReject> ON | OFF
*RST: OFF
Usage: Asynchronous command
Firmware/software: FW 2.25
Trigger