Remote control commands
R&S
®
RTE
1258User Manual 1326.1032.02 ─ 20
<MeasType>
parameter value
(remote control)
Measurement
(manual opera-
tion)
Description, result Suffix
<n>
FREQuency Frequency Frequency of the signal. The result is based on the
period measurement.
18
PDCYcle Pos. duty cycle Positive duty cycle. The measurement requires at least
one complete period of a triggered signal.
19
NDCYcle Neg. duty cycle Negative duty cycle. The measurement requires at least
one complete period of a triggered signal.
20
CYCarea Cycle area Area (integral) beneath one cycle 21
CYCMean Cycle mean Mean value of one cycle 22
CYCRms Cycle RMS The RMS (Root Mean Square) value of one cycle 23
CYCStddev Cycle σ (S-dev) Standard deviation of one cycle 24
PULCnt Pulse count Number of positive or negative pulses of the waveform,
or both
25
DELay Delay Time difference between any edges of two measurement
sources at any reference level. The measurement result
is negative if the edge of the second source comes
before the edge of the first source.
26
PHASe Phase Phase difference between two waveforms 27
BWIDth Burst width Duration of one burst, measured from the first edge to
the last
28
PSWitching Pos. switching Settling time at rising edges 29
NSWitching Neg. switching Settling time at falling edges 30
PULSetrain Pulse train Duration of N positive pulses, measured from the rising
edge of the first pulse to the falling edge of the N-th
pulse. N has to be configured.
31
EDGecount Edge count Number of positive or negative edges of the waveform,
or both
32
SETup Setup time Parameters to query the setup and hold times.
Use these parameters only in following queries:
MEASurement<m>:ARES?
MEASurement<n>:RESult:.. commands
33
HOLD Hold time
SHT Setup/Hold time Setting parameter to enable Setup/Hold time measure-
ments.
Use this parameter only as setting in:
MEASurement<m>:MAIN on page 1253
35
SHR Setup/Hold ratio Setup/Hold ratio measurement. Setup/Hold ratio is the
ratio of the setup time to the sum of hold and setup time:
T
Setup
/
(T
Setup
+
T
Hold
)
Use this parameter as setting to activate the Setup/Hold
ratio measurement in:
MEASurement<m>:MAIN on page 1253
It is also used in the following queries:
MEASurement<m>:ARES?
MEASurement<n>:RESult:..
Used also in queries with and commands.
36
Automatic measurements

Table of Contents