Remote control commands
R&S
®
RTE
1263User Manual 1326.1032.02 ─ 20
MEASurement<m>:AMPTime:PTCount <PulseCount>
Sets the number of positive pulses for the pulse train measurement. It measures the
duration of N positive pulses from the rising edge of the first pulse to the falling edge of
the N-th pulse.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<PulseCount> Range: 1 to 2147483647
Increment: 1
*RST: 1
Usage: Asynchronous command
MEASurement<m>:AMPTime:ESLope
<EdgesSlope>
Sets the edge direction to be counted: rising edges, falling edges, or both. The setting
is only relevant for edge count measurement (MEASurement<m>:MAIN EDGecount
or MEASurement<m>:ADDitional EDGecount,ON).
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<EdgesSlope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
MEASurement<m>:AMPTime:CSLope <SetHoldClkSlp>
Sets the edge of the clock from which the setup and hold times are measured.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<SetHoldClkSlp> POSitive | NEGative | EITHer
EITHer
The clock edges next to the data edge are considered regard-
less of the clock slope.
*RST: POSitive
Usage: Asynchronous command