Remote control commands
R&S
®
RTE
1312User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 111, 115 to 118
Source waveform of the measurement, see Chapter 18.4.1,
"Waveform suffix", on page 1072.
Parameters:
<Value>
Example:
REFLevel2:ABSolute:TDIStance 0.0002
Sets the top distance for Ch1 to 0.2 mV. C1W1 corresponds to
suffix number 2.
Example: See: "Manual reference level definition using absolute values"
on page 1058
Usage: Asynchronous command
REFLevel<m>:ABSolute:BDIStance <Value>
The distance between the lower reference level and the low signal value.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 111, 115 to 118
Source waveform of the measurement, see Chapter 18.4.1,
"Waveform suffix", on page 1072.
Parameters:
<Value>
Example:
REFLevel2:ABSolute:BDIStance 0.0002
Sets the bottom distance for Ch1 to 0.2 mV. C1W1 corresponds
to suffix number 2.
Example: See: "Manual reference level definition using absolute values"
on page 1058
Usage: Asynchronous command
REFLevel<m>:ABSolute:MLEVel <Value>
The middle level between high and low signal level. The value is adjusted automati-
cally if you change the high or low signal levels. Vice versa, if you change the middle
level, the high and low signal levels are adjusted.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 111, 115 to 118
Source waveform of the measurement, see Chapter 18.4.1,
"Waveform suffix", on page 1072.
Parameters:
<Value>
Example:
REFLevel2:ABSolute:MLEVel 0.05
Sets the middle signal level for Ch1 to 50 mV. C1W1 corre-
sponds to suffix number 2.
Automatic measurements