Remote control commands
R&S
®
RTE
1339User Manual 1326.1032.02 ─ 20
<MaskSegmIdx> Number of the segment. Counting starts from 0.
Usage: Query only
MTESt:SEGMent:POINt:X <MaskTestName>, <MaskSegmIdx>,
<MaskSegmPtIdx>,<X>
MTESt:SEGMent:POINt:X? <MaskTestName>, <MaskSegmIdx>, <MaskSegmPtIdx>
Defines the x-value of the mask segment point.
Parameters:
<X> Range: -100E+24 to 100E+24
Increment: 1E-6
*RST: 0
Default unit: s
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
<MaskSegmIdx> Number of the segment. Counting starts from 0.
<MaskSegmPtIdx> Number of the point. Counting starts from 0.
Example: See Chapter 18.3.3.1, "Creating a user mask", on page 1060
Usage: Asynchronous command
MTESt:SEGMent:POINt:Y <MaskTestName>, <MaskSegmIdx>,
<MaskSegmPtIdx>,<Y>
MTESt:SEGMent:POINt:Y? <MaskTestName>, <MaskSegmIdx>, <MaskSegmPtIdx>
Defines the y-value of the mask segment point.
Parameters:
<Y> Range: -100E+24 to 100E+24
Increment: 1E-6
*RST: 0
Default unit: V
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
<MaskSegmIdx> Number of the segment. Counting starts from 0.
<MaskSegmPtIdx> Number of the point. Counting starts from 0.
Example: See Chapter 18.3.3.1, "Creating a user mask", on page 1060
Usage: Asynchronous command
MTESt:SEGMent:RESCale:RECalculate <MaskTestName>, <MaskSegmIdx>
Multiplies and adds the given x- and y-factors and offsets to the coordinates of all
points of the selected mask segment.
Mask testing